The chamber will not pump down. What’s the reason ?

There are three possible causes:
(1) The leakage from the Ferrofluidic Feedthroughs’ failure;
(2) The leakage from the static seal(such as O-ring, gasket) for the mounting Ferrofluidic Feedthroughs;
(3) The leakage from some other places of the equipment.

The startup/running torque of the Ferrofluidic Feedthroughs is too big. What’s the reason ?

There are five possible causes:
(1) The Ferrofluidic Feedthroughs has been left in damp environment without rotation for a long time;
(2) The environment temperature is very low;
(3) There are many particles coming from chemical process have deposited on the seal gap of the Ferrofluidic Seal;
(4) The bearing grease has deteriorated;
(5) The bearing is damaged.

The Ferrofluidic Feedthroughs cannot rotate.

Please return the failed product to us which need to be overhauled by Vigortec. Do not disassemble it by yourself.

The Ferrofluidic Feedthroughs are invaded with water/organic solvents directly.

Please return the failed product to us which need to be overhauled by Vigortec. Do not disassemble it by yourself.

Ferrofluidic Feedthroughs are dropped or striked accidentally.

Please return the failed product to us which need a detailed inspection for its dimensions and performance by Vigortec. Do not disassemble it by yourself.
Found black/brown liquid spilled from the Ferrofluidic Feedthroughs.
Stop the equipment immediately and return it to Vigortec for overhaul.

The Ferrofluidic Feedthroughs rotate with an abnormal noise.

Stop the equipment immediately and return it to Vigortec for overhaul.